• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FB INSERT SZ 4 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FB INSERT SZ 4 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-20-407
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Adhesion(s) (1695)
Date of Event 07/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical adverse event received for arthrofibrosis.Event is not serious and is considered moderate.Event is not related to device and is related to procedure.Date of implantation: (b)(6) 2021.Date of event (onset): (b)(6) 2021.(left knee).Treatment: knee manipulation under anesthesia on (b)(6) 2021.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSERT SZ 4 7MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key12573887
Report Number1818910-2021-21892
Device Sequence Number19810446
Product Code JWH
UDI-Device Identifier10603295045540
UDI-Public10603295045540
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1516-20-407
Device Catalogue Number151620407
Device Lot NumberJA8685
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/21/2021
Supplement Date Received by Manufacturer10/12/2021
Initial Report FDA Received Date10/04/2021
Supplement Report FDA Received Date10/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSERT SZ 4 7MM; ATTUNE CR FEM LT SZ 4 NAR CEM; ATTUNE FB TIB BASE SZ 3 CEM; ATTUNE MEDIAL ANAT PAT 35MM; ATTUNE CR FB INSERT SZ 4 7MM; ATTUNE CR FEM LT SZ 4 NAR CEM; ATTUNE FB TIB BASE SZ 3 CEM; ATTUNE MEDIAL ANAT PAT 35MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient Weight96
-
-