| Model Number |
55-0514 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646)
|
| Date of Event |
09/21/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient had a hip replacement done at (b)(6) hospital on the (b)(6) 2016 and received a s-rom hip.The patient has been complaining of increasing groin and thigh pain for some time.Following further investigation, surgeon decided the stem was undersized and was now loose.The loose s-rom was removed today and the patient was revised to an exeter cemented stem (stryker).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : a worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received states that both stem and sleeve were loose - undersized.The affected side of the patient was the right hip.
|
| |
|
Search Alerts/Recalls
|