| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Foreign: (b)(6).The event was confirmed with product received.Visual inspection found the seal width at the top of the pouch to be inconsistent.Dimensional analysis found one measurement to fall below the minimum of 6.0 mm.Device history record was reviewed and no discrepancies relevant to the reported event were found.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.A corrective action was opened to further investigate the reported event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported by the distributorship that the products were found to be nonconforming.During incoming inspection, the team member found the seal width was less than 1/4 inch (6.35mm).No further event information available at the time of this report.
|
| |
|