| Model Number |
7212 |
| Medical Device Problem Codes |
Contamination (1120); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon came out of the package occurred.A 3.00mm x 12mm nc emerge balloon catheter was prepared, however, it was observed that the packaging was damaged and balloon came out of the inner package with the tip poking out.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon came out of the package occurred.A 3.00mm x 12mm nc emerge balloon catheter was prepared, however, it was observed that the packaging was damaged and balloon came out of the inner package with the tip poking out.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of an nc emerge balloon catheter and its originally shipped packaging and outer box.The returned outer box was visually inspected with no damage present past opening of the box.The product pouch was visually inspected.The product pouch contained a unit for batch 27144786 which matched the returned packaging.The tyvek seal was visually inspected.The bottom tyvek (white) seal was pulled and curled up on the back flap of the packaging far enough that the device was visibly capable of being exposed out of the package.However, all seals on the package were complete and sealed at one point without error.Operations engineering was contacted for verification.Inspection of the remainder of the device, apart form the observed damage to packaging, revealed no other damage or irregularities.The complaint of the device having packaging damage was confirmed.Photo media reviewed matched analysis of device.
|
| |
|
Search Alerts/Recalls
|