• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; STEM EXTENSION, CEMENTED, 15X30MM, STRAIGHT, BULLET-TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; STEM EXTENSION, CEMENTED, 15X30MM, STRAIGHT, BULLET-TIP Back to Search Results
Model Number KSC01530E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event was reported to onkos surgical on (b)(6) 2021 with limited information.Additional information was received on 28 september 2021 regarding this adverse event which identified the suspect medical device captured in this report as being implanted at the time of the alleged infection.The device will not be returned for investigation.The investigation is in progress.When the investigation is complete, a supplemental report will be submitted accordingly.Multiple mdrs were reported for this event: #3013450937-2021-00196, #3013450937-2021-00197, #3013450937-2021-00198, #3013450937-2021-00199, #3013450937-2021-00200, #3013450937-2021-00223, #3013450937-2021-00224, #3013450937-2021-00225, #3013450937-2021-00227.
 
Event or Problem Description
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.During the revision surgery, the following implants were revised: eleos distal femur axial pin, eleos distal femur, eleos tibial poly spacer, eleos segmental stem, and eleos tibial hinge component.Additional information was received that identified the following implants remain implanted from the patient's original eleos surgery: tibial baseplate, 2 tibial block augments, stem extension, and modular tibial base stem cap.Additional information regarding this event has not been provided.
 
Event or Problem Description
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.During the revision surgery, the following implants were revised: eleos distal femur axial pin, eleos distal femur, eleos tibial poly spacer, eleos segmental stem, and eleos tibial hinge component.Additional information was received that identified the following implants remain implanted from the patient's original eleos surgery: tibial baseplate, 2 tibial block augments, stem extension, and modular tibial base stem cap.Additional information regarding this event has not been provided.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that could have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: h3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
STEM EXTENSION, CEMENTED, 15X30MM, STRAIGHT, BULLET-TIP
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
MDR Report Key12576705
Report Number3013450937-2021-00226
Device Sequence Number3920923
Product Code KRO
UDI-Device IdentifierB278KSC01530E0
UDI-Public+B278KSC01530E0
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberKSC01530E
Device Catalogue NumberKSC01530E
Device Lot Number1859503
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2021
Supplement Date Received by Manufacturer11/04/2021
Initial Report FDA Received Date10/05/2021
Supplement Report FDA Received Date11/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000007E, ELEOS DISTAL FEMUR; P/N 25001212E, ELEOS TIBIAL POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N 25002203E, ELEOS TIBIAL BASEPLATE; P/N CS-17120-03M, ELEOS SEGMENTAL STEM; P/N KSC01500E, ELEOS MODULAR TIBIAL BASE STEM CAP; P/N KTAGB305E, ELEOS TIBIAL BLOCK AUGMENT; P/N KTAGB305E, ELEOS TIBIAL BLOCK AUGMENT; P/N THSMWOS01M, ELEOS TIBIAL HINGE COMPONENT; P/N 25000007E, ELEOS DISTAL FEMUR; P/N 25001212E, ELEOS TIBIAL POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N 25002203E, ELEOS TIBIAL BASEPLATE; P/N CS-17120-03M, ELEOS SEGMENTAL STEM; P/N KSC01500E, ELEOS MODULAR TIBIAL BASE STEM CAP; P/N KTAGB305E, ELEOS TIBIAL BLOCK AUGMENT; P/N KTAGB305E, ELEOS TIBIAL BLOCK AUGMENT; P/N THSMWOS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
-
-