BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7213
Device Problems Entrapment of Device (1212); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2021
Event Type  malfunction  
Event Description
Reportable based on additional information on 15-sept-2021.It was reported that catheter entrapment occurred.The 85% in-stent restenosis (isr) target lesion was located in the non-calcified and non-tortuous proximal saphenous vein graft to obtuse marginal.A 4.50mm x 15mm nc emerge balloon catheter was advanced for dilatation and was inflated twice at 26 atmospheres for 20 seconds.The device was fully deflated, however, it would not come out of the guide catheter.The physician then pulled everything out together and the procedure was completed successfully.No patient complications were reported.It was further reported that the balloon was also stuck on the wire.The physician had to remove the guide, balloon, and wire together.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12576716
MDR Text Key274745950
Report Number2134265-2021-11922
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729846574
UDI-Public08714729846574
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141236
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/18/2023
Device Model Number7213
Device Catalogue Number7213
Device Lot Number0027332212
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/15/2021
Initial Date FDA Received10/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-