|
This complaint was reopened in reference to (b)(4).Investigation: ( description incoming product condition) the valve was received inserted in an unknown liquid.The device was noted to have been manufactured in april 2016.At the time of submission the valve was set to pressure range 3 cmh2o.Summary: first we performed a visual inspection with the valve.Apart from slight scratches on the housing, we could not detect any deformations or other abnormalities.To investigate if maybe a blockage have caused the assumed problems with re-programming the valve, we performed a permeability test.The test results that the valve was penetrable.For further information we tried to adjust the valve from 0 cmh2o up to 20 cmh2o and down again in the same way in steps of 2 cmh2o.Based on our investigation results we cannot confirm any difficulties in adjusting the valve.It was possible to adjust the valve in all pressure ranges.The measurement of the braking force could additionally confirm that the measured value was within the accepted tolerance.As a final examination we carried out a computer controlled test.Here the progav 2.0 was tested in the lying position with a pressure range of 5 cmh2o.Normally we expected a pressure of 5 cmh2o having an opening pressure of 5 cmh2o.The opening pressure at the reference flow level of 5 ml/h is measured 4,08 cmh2o.The progav 2.0 operates reliably within the accepted tolerance (acc.Tolerance 5 +/- 2 cmh2o).In order to verify whether the valve examined here has been influenced by the known risks of hydrocephalus therapy, as for example by natural substances (protein, blood or tissue particles) in the cerebrospinal fluid, we have finally opened the valve.After the valve has been opened we have found obvious deposits inside the valve, which could have impaired the product performance in the past.Based on our investigation results we cannot confirm that it was not possible to adjust the valve in lower values.The valve could easily be adjusted in all pressure ranges.However, we are able to detect deposits inside the valve that might have been a cause in the past.Based on our investigation results we could not detect any failure with the valve at the time of delivery.No further actions required.
|