(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.(b)(4).Investigation summary: fifteen packets of product code j259h were returned to ethicon inc for analysis with the packaging closed.Upon initial inspection to the samples no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.Functional test was performed, and the tensile strength result were above the minimum requirements.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.What is the procedure date? could you please clarify what do you mean by "the happened multiple times"? could you please confirm if the other issues occurred mention in ed was reported in other pcs numbers? if yes, could you please provide pc numbers? could you please confirm if medical intervention was needed? removed and resutured? any medical treatment provided, were prescription steroids administered, antibiotics? if yes, please provide medicine name, strength and dose.The single complaint was reported with multiple events.There are no additional details regarding the additional patient events.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Note: events reported in 2210968-2021-09217.
|