BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR101209
Device Problems Break (1069); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event Description
It was reported that, during preventative maintenence, they performed a drill test on two anspach emax 2 plus hand piece - rohs and both started to smoke after 10 seconds and were too hot to touch.Also, surgeons have stated that the motor sound.It is suspected that internal lubricant has diminished and bearings not working correctly.No case involved; therefore, there was no patient involvement.
 
Manufacturer Narrative
H3, h6: the anspach emax 2 plus hand piece - rohs, part number pfsr101209, serial sn (b)(4) and intended to be used for treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was confirmed.A drill test was performed and a louder-than-expected noise was coming from the drill as well as smoke.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most probable cause of the reported problem is motor failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Manufacturer Narrative
H3, h6: the anspach emax 2 plus hand piece - rohs, part number pfsr101209, serial sn (21)(b)(6) and intended to be used for treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was confirmed.A drill test was performed and a louder-than-expected noise was coming from the drill as well as smoke.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most probable cause of the reported problem is motor failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANSPACH EMAX 2 PLUS HAND PIECE - ROHS
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12578999
MDR Text Key277717015
Report Number3010266064-2021-00706
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR101209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ROB00049-NAVIO SURGICAL SYSTEM UK
-
-