• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/18; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/18; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364491
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment.The balloon could not be inflated, and therefore the stent could not be deployed.
 
Additional Manufacturer Narrative
Combination product: yes the returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the balloon has not been inflated but both balloon shoulders are slightly unfolded.The stent is also slightly opened at both ends.Microscopic inspection revealed a constriction of the outer shaft at the proximal balloon weld.The findings indicate that the shaft deformations were most likely caused by a tensile force which has been applied to the distal part of the instrument.A reference transportation wire and a reference guidewire could be introduced into the guidewire lumen of the instrument.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is inflated and tested for air tightness by means of a helium leakage test.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause for the complaint event is most likely related to the handling during the preparations for the intervention.It should be noted that, according to the ifu, the user is advised to exercise care during device handling to reduce the possibility of accidental breakage, bending, kinking of the stent system shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.5/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key12579830
Report Number1028232-2021-05480
Device Sequence Number19810267
Product Code NIQ
UDI-Device Identifier07640130439061
UDI-Public7640130439061
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2022
Device Model Number364491
Device Catalogue NumberSEE MODEL NO.
Device Lot Number06202456
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2021
Initial Date Received by Manufacturer 10/04/2021
Supplement Date Received by Manufacturer10/27/2021
Initial Report FDA Received Date10/05/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-