• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HUDSON ET TUBE, HVT, 8.0; TUBE, TRACHEAL (W/WO CONNECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUDSON ET TUBE, HVT, 8.0; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Model Number IPN048897
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A verification of failure mode reported in the current manufacturing process was conducted as follows: 315 samples were taken from the current production, the samples were inspected according to qip, the revised characteristics comply with the requirement.Defect reported "torn/ruptured - cuff" was not observed in the current manufacturing process.A device history record review could not be conducted since the lot number of the device was not provided.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported "cuff of the endotracheal tube was blown completely causing clinicians to have to reintubate the patient".No patient harm or injury reported.Patient condition reported as "fine".
 
Event or Problem Description
It was reported "cuff of the endotracheal tube was blown completely causing clinicians to have to reintubate the patient".No patient harm or injury reported.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned one unit 5-10316 et tube, hvt, 8.0 for investigation.The returned et tube was visually examined with and without magnification.Visual examination of the returned et tube revealed that the sample appears used as there is biological material present on the device.A functional inspection was performed to attempt to inflate and deflate the balloon cuff on the returned et tube.A lab inventory syringe was filled with air and attached to the valve of the pilot balloon.Upon injection of air, both the pilot balloon and cuff immediately inflated.The syringe was then removed and the cuff and pilot balloon were inspected for leaks.No leaks were detected.The syringe was then reattached in order to deflate the balloons.Both the pilot balloon and balloon cuff were able to deflate.The et tube was submerged underwater and an attempt to inflate the balloon cuff was made in order to detect any leaks.Upon injection of air, no leaks were detected.The cuff was left inflated for approximately 30 minutes with no leaks observed.No functional issues were found with the returned sample.The ifu for this product states, "the cuff inflation system (cuffs, pilot balloons, valves) should be checked by inflation and deflation prior to use.If cuff is damaged, the tube should not be used." "care should be taken to avoid damaging the cuffs during intubation.If any one of the cuffs is damaged, the tube should not be used." the ifu also states, "deflate all cuffs prior to repositioning the tube.Movement of the tube with cuffs inflated could result in patient injury or in damage to the cuffs, requiring a tube change.Verify the position of the tube after each repositioning." the reported complaint could not be confirmed based upon the sample received.The returned et tube was able to inflate and deflate with no difficulty.All et tubes are 100% inspected for both inflation and deflation at the time of manufacturing.No functional issues were found with the returned device.Teleflex will continue to monitor and trend on reports of this nature.Corrected data: section h.6.-investigation type code '3331' removed as it was inadvertently put on the form.
 
Event or Problem Description
It was reported "cuff of the endotracheal tube was blown completely causing clinicians to have to reintubate the patient".No patient harm or injury reported.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned one unit 5-10316 et tube, hvt, 8.0 for investigation.The returned et tube was visually examined with and without magnification.Visual examination of the returned et tube revealed that the sample appears used as there is biological material present on the device.A functional inspection was performed to attempt to inflate and deflate the balloon cuff on the returned et tube.A lab inventory syringe was filled with air and attached to the valve of the pilot balloon.Upon injection of air, both the pilot balloon and cuff immediately inflated.The syringe was then removed and the cuff and pilot balloon were inspected for leaks.No leaks were detected.The syringe was then reattached in order to deflate the balloons.Both the pilot balloon and balloon cuff were able to deflate.The et tube was submerged underwater and an attempt to inflate the balloon cuff was made in order to detect any leaks.Upon injection of air, no leaks were detected.The cuff was left inflated for approximately 30 minutes with no leaks observed.No functional issues were found with the returned sample.The ifu for this product states, "the cuff inflation system (cuffs, pilot balloons, valves) should be checked by inflation and deflation prior to use.If cuff is damaged, the tube should not be used." "care should be taken to avoid damaging the cuffs during intubation.If any one of the cuffs is damaged, the tube should not be used." the ifu also states, "deflate all cuffs prior to repositioning the tube.Movement of the tube with cuffs inflated could result in patient injury or in damage to the cuffs, requiring a tube change.Verify the position of the tube after each repositioning." the reported complaint could not be confirmed based upon the sample received.The returned et tube was able to inflate and deflate with no difficulty.All et tubes are 100% inspected for both inflation and deflation at the time of manufacturing.No functional issues were found with the returned device.Teleflex will continue to monitor and trend on reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON ET TUBE, HVT, 8.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
MDR Report Key12579881
Report Number3003898360-2021-00889
Device Sequence Number19705098
Product Code BTR
UDI-Device Identifier14026704616760
UDI-Public14026704616760
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K822082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN048897
Device Catalogue Number5-10316
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2021
Initial Date Received by Manufacturer 09/10/2021
Supplement Date Received by Manufacturer10/29/2021
10/29/2021
Initial Report FDA Received Date10/05/2021
Supplement Report FDA Received Date11/01/2021
11/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
-
-