| Model Number |
EVPMP |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned for evaluation.When the product evaluation has been completed, a supplemental report will be submitted.The device history record review was completed and all inspections passed with no non-comformances.
|
| |
|
Event or Problem Description
|
|
The pump unit failed functional test for pressure at various flows with the difference pressure outlet ad at 150 normal liters per hour.The measurement is 136.3 mmhg and the limits are from plus or minus 10 mmhg.The test technician found that the pumps manifold bottom had cracks around the pems for the manifold lid.All tests passed after the manifold bottom was replaced.There were no error messages or alarms that displayed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
New information was received that the test technician reported the incorrect information for section v whether erroneous hemodynamic or pressure values were observed regarding the in process failure of the evpmp during lab testing.There were no erroneous values noted and this was due to it getting the failing values from test equipment when not connected to an ev1000 monitor.This requires a correction of mdr reportability to not reportable because although the pump unit failed testing there were no erroneous values reported.
|
| |
|