• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU150
Medical Device Problem Codes Break (1069); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age, weight and ethnicity: unknown, not provided.If implanted; give date: n/a.(not applicable).The lens was not implanted and a back up lens was used.If explanted; give date: n/a (not applicable).The lens was not implanted so not explanted.Device evaluation: product evaluation was not performed because the product has not been returned.The complaint issue reported could not be verified and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the tecnis eyhance preloaded intraocular lens (iol) inserter system seemed to crack when turned and the iol felt rock hard and cracked even after hydration.The surgeon was unable to implant into the patient's eye and had to use a competitive product.No patient injury was reported.No additional information was provided.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: 11/4/2021.Device evaluation: the complaint simplicity and lens was received in plastic bag.Visual inspection under magnification revealed viscoelastic residue on the haptics and optic body of the complaint lens.Further inspection revealed a detached haptic and lens damage.A detached haptic was received with the iol that belonged to a different lens.Inspection under magnification, of the cartridge revealed, trace amounts of viscoelastic residue inside of the cartridge.The cartridge was observed to have a crack on the tube and the cartridge tip.The cartridge was also observed to be damaged (bent).The complaint issue lens damage was confirmed; however, this may be attributed to using an inadequate amount of ovd, suggested by the trace amounts found in the cartridge.The complaint issue device advancement issue could not be confirmed.Therefore, no product deficiency could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
1700 e st andrew place
santa ana, CA 92705
7142478552
MDR Report Key12580674
Report Number2020664-2021-07632
Device Sequence Number3376020
Product Code MJP
UDI-Device Identifier05050474746565
UDI-Public(01)05050474746565(17)240627
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2024
Device Model NumberDIU150
Device Catalogue NumberDIU150U210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2021
Supplement Date Received by Manufacturer12/09/2021
Initial Report FDA Received Date10/05/2021
Supplement Report FDA Received Date01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-