Investigation was performed based on the device log file and the additionally provided information.A dräger technician was on-site and confirmed the reported restarts.A mechanically damaged 28v dc cable on the control board was identified to be root cause for the reported symptom.Reportedly, the cable was damaged by the biomed during replacement of the control box, which was performed before the event.A picture of the affected area was provided.Based on that, it could be confirmed that by this damage, the cable from power supply to the dc supply was affected leading to the reported symptom.Also based on the log analysis, this damaged dc cable as root cause could be strengthened.Even though the log records were overwritten and only started at 11:17am on the reported date of event, entries were found pointing to a failure pattern correlating with the identified cable issue.The power supply itself was working but did not provide enough energy to fully operate the fabius.In consequence, energy was taken simultaneously from the backup batteries as well until the backup batteries were depleted and the fabius finally shut down.After the shutdown, the fabius mri was connected to mains power in switch-on state.The power supply provided sufficient energy to start the device but when more energy was required, it again shut down and restarted.This sequence has been repeated several times and could also be comprehended based on the log analysis as a high number of "system power on"-messages was found.As only the "switching on" sequence is logged, the device had to be switched off in between.This fits to the reported symptom of multiple restarts, so that the fabius device was switching on and off all the time.In general, a power loss is accompanied by a corresponding alarm buzzer whereby the fabius mri additionally indicates this state by a warning light.In case the capacity of the backup battery falls below 20% respectively 10%, corresponding alarms are posted.Dräger finally concludes that the damaged dc power supply cable was the root cause for the reported symptom.It can be concluded that the replacement of the control box prior the case in question by the biomed has damaged the cable and finally resulted in the reported symptoms.This case can be considered as single event as no similar case has been reported before.
|