DEPUY INTERNATIONAL LTD - 8010379 C-STEM AMT SZ1 STD OFFSET; C-STEM HIP IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 1570-04-070
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 09/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
I attended surgery today for a revision hip case where a c-stem amt primary was replaced (due to stem fracture) with a reclaim.This was scheduled revision surgery so no implications for extended surgery time.Primary surgery details - surgery date: (b)(6) 2015; revision surgery details - surgery date: (b)(6) 2021.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) investigation summary: no device associated with this report was received for examination.Visual examination of the provided x-ray image confirmed the reported event.The femoral stem was fractured.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: a worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C-STEM AMT SZ1 STD OFFSET
Common Device Name
C-STEM HIP IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12582624
Report Number1818910-2021-22068
Device Sequence Number14296789
Product Code KWA
UDI-Device Identifier10603295059592
UDI-Public10603295059592
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K082239
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Model Number1570-04-070
Device Catalogue Number157004070
Device Lot NumberD15101931
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2021
Supplement Date Received by Manufacturer12/06/2021
Initial Report FDA Received Date10/06/2021
Supplement Report FDA Received Date12/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 28 +12 BLK; C STEM VOID CENTALIZER 12 DM; C-STEM AMT SZ1 STD OFFSET; DEPUY CMW 2G 40G; MARATHON XLPE CEMENT CUP 28/45; MARATHON XLPE CEMENT CUP 28/45; ARTICUL/EZE BALL 28 +12 BLK; C STEM VOID CENTALIZER 12 DM; C-STEM AMT SZ1 STD OFFSET; DEPUY CMW 2G 40G; MARATHON XLPE CEMENT CUP 28/45; MARATHON XLPE CEMENT CUP 28/45
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-