|
Additional information was received on sep 15, 2023 and section h11 should be reported as: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging a continuous positive airway pressure (cpap) device's sound abatement foam became degraded and caused the patient to develop lung nodules.The medical intervention that the patient received in response to the event is currently unknown there was no report of serious or permanent harm or injury.The device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was no mention of visual findings to the external part of the device.The device's downloaded logs were reviewed.There were four errors found.The manufacturer concludes there was no evidence of sound abatement foam degradation in the device.
|