Model Number Z148H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
 
Event or Problem Description
It was reported that a patient underwent a cleft lip and palate procedure on (b)(6) 2021 and suture was used.During the procedure, the needle pulled off of the suture whiling sewing upper lip at the second stitch.Changed another one to complete the surgery.No adverse patient consequences were reported.No additional information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/02/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 evaluation: h3 investigation summary: one labeled winding former a with a needle and a suture tangled around the tray of product code z148 were returned for analysis.In order to evaluate the conditions of the returned sample, the swage area and hole of the needle were noted with marks that appears to be by surgical instrument and the hole was observed with suture remnant.To avoid this kind of damage: grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.During the visual inspection of the suture, body fluids and marks on the surface due to use were found, the end was cut by sharp edge.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.Functional test was not performed, due to needle was received detached from suture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The condition of the sample received indicates improper handling of the device.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII VIO 30IN 5-0 D/A RB-2
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
MDR Report Key12584852
Report Number2210968-2021-09265
Device Sequence Number14430142
Product Code NEW
UDI-Device Identifier10705031060272
UDI-Public10705031060272
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberZ148H
Device Catalogue NumberZ148H
Device Lot NumberRCMLZL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2021
Initial Date Received by Manufacturer 09/09/2021
Supplement Date Received by Manufacturer10/07/2021
Initial Report FDA Received Date10/06/2021
Supplement Report FDA Received Date11/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age5 YR
-
-