| Catalog Number |
705.13P |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A product sample has been received by the manufacturer and it is awaiting evaluation.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported the forceps tip stopped opening during a vitrectomy procedure.The forceps was replaced with another one and the procedure was completed.There was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Sample received in opened original packaging including cover foil.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The received forceps sample was visually inspected with the aid of a photomicroscope with various magnifications.After cleaning, it was found that the tube is loose which block the forceps from activating.The customer¿s complaint was confirmed.Root cause was unable to verify.Why the tube was loose and blocked the forceps from activating could not be determined.This device passed the 100% control and was conform when leaving the production.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|