| Model Number |
68LP-550 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, left side deflation.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified, opening, wear abrasion, crease fold.Leak test was performed, and found opening on anterior.A microscopic analysis was performed which identify, crease smooth opening on anterior.The fill test inspection was performed, the result is no blockage.Based on the device analysis, the final assessment is: a crease smooth opening on anterior assessed, to a fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|