• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 68LP-550
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Healthcare professional reported left side deflation.Device was explanted.
 
Event or Problem Description
Healthcare professional reported, left side deflation.Device was explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified, opening, wear abrasion, crease fold.Leak test was performed, and found opening on anterior.A microscopic analysis was performed which identify, crease smooth opening on anterior.The fill test inspection was performed, the result is no blockage.Based on the device analysis, the final assessment is: a crease smooth opening on anterior assessed, to a fold flaw opening.
 
Event or Problem Description
Healthcare professional reported left side deflation.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key12586608
Report Number9617229-2021-51282
Device Sequence Number16497790
Product Code FWM
UDI-Device Identifier10888628000667
UDI-Public10888628000667
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/25/2021
Device Model Number68LP-550
Device Catalogue Number68LP-550
Device Lot Number3049368
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/07/2021
Supplement Date Received by Manufacturer10/26/2021
11/29/2021
Initial Report FDA Received Date10/06/2021
Supplement Report FDA Received Date11/24/2021
12/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
Patient SexFemale
Patient Weight89 KG
Patient EthnicityHispanic
-
-