Model Number 3660 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 09/17/2021 |
Event Type
Injury
|
Event Description
|
It was reported the patient had an unrelated procedure where it is unknown if the device was set to surgery mode or not.In turn, the ipg became inoperable.Troubleshooting is pending.
|
|
Manufacturer Narrative
|
The implant was not returned for analysis.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
|
|
Event Description
|
Additional information received indicates that the patient underwent surgical intervention during which the ipg was explanted and replaced with no complications.
|
|