Model Number X060-1520
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A visual assessment of the returned complaint inserter/compressor showed an instrument with minimal use, as identified by crisp laser markings and a lack of surface scratches.The pin of the instrument distal tip was fractured where it connects with the instrument handle.A functionality assessment was not performed due to the damaged condition of the returned inserter/compressor, which was removed from distributable inventory.It may be possible that application of excessive compression force may have contributed to one of the instrument distal tips fracturing from the instrument.The instrument tip is secured into the instrument handle with a cylinder that has minor movement to allow for implant engagement and patient anatomy.It may be possible that plate angulation due to patient anatomy or over compressing the instrument may concentrate applied force to the instrument distal tip, resulting in an instrument malfunction.
 
Event or Problem Description
The company received notification on (b)(6) 2021 of a system inserter/compressor that malfunctioned during a surgical procedure.It was reported that one of the distal tips of the instrument fractured while compressing interspinous plates after being placed.There were no known patient complications associated with this complaint.Alternate available instruments were used to successfully complete the surgical procedure.The complaint instrument arrived at the manufacturer for assessment on 9/28/2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXLE INTERSPINOUS FUSION SYSTEM
Common Device Name
SPINOUS PROCESS PLATE
Manufacturer (Section D)
XTANT MEDICAL
664 cruiser lane
belgrade MT 59714
MDR Report Key12586780
Report Number3005031160-2021-00021
Device Sequence Number3067699
Product Code PEK
UDI-Device IdentifierM697X06015201
UDI-PublicM697X06015201
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K130438
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX060-1520
Device Lot Number21179
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2021
Initial Report FDA Received Date10/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient Weight104
-
-