• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SMHX700
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Pocket Erosion (2013)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: implant exposure/capsular contracture (b)(4).
 
Event or Problem Description
It was reported that a caucasian female who had 700cc mentor memorygel breast implants placed experienced right sided capsular contracture, right sided implant exposure, and the presence of a hypertrophic scar post procedure.As a result, explant was performed on (b)(6) 2021.See 1645337-2021-11187 for contralateral prosthesis report.
 
Additional Manufacturer Narrative
The impacted product was received by the failure analysis lab on october 7, 2021.The investigation of the returned device was completed by the failure analysis lab on october 13, 2021.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During the visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Extrusion may occur when the wound has not closed or when tissue covering the implants weakens.The incidence of extrusion has been shown to increase when undue pressure is applied on the tissue located over the device, trauma to surrounding tissues may lead to thrombosis, delayed wound healing, improper size, placement, larger sized implants, use of steroids in the surgical pocket and microwave diathermy.Radiation therapy has been reported to increase the likelihood of extrusion.Extrusion may require additional surgery and possible removal of the implant, which may result in additional scarring and/or loss of breast tissue.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
MDR Report Key12587305
Report Number1645337-2021-11188
Device Sequence Number4584339
Product Code FTR
UDI-Device Identifier10081317027925
UDI-Public10081317027925
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/18/2025
Device Model NumberSMHX700
Device Catalogue NumberSMHX700
Device Lot Number9525983
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2021
Initial Date Received by Manufacturer 09/15/2021
Supplement Date Received by Manufacturer10/07/2021
Initial Report FDA Received Date10/06/2021
Supplement Report FDA Received Date10/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-