• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC CAPSTONE CONTROL¿ SPINAL SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC CAPSTONE CONTROL¿ SPINAL SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 3990003
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information received from healthcare provider via manufacturer representative regarding an event happened during intra-op for a patient undergoing l2-s2ai-psf, l4/5-plif.The pre-operative diagnosis was mentioned as lumbar spinal canal stenosis.It was reported that, the short inserter was found to have a crack at thread and the other inserter shafts were not attached to implant due to thread damage.The inserter tip was deformed.The inserter shaft was stripped at the tip and scratches and burrs were found at threads on handle side.The other inserter had scratches at tip threads and scratches and stripping found at threads.The was a delay of less than 60 minutes reported as a result of this event.Patient medical history includes ht, dm, cerebral infarction, two-sided tka, dvt.There was no health damage in patient reported as a result of this event.No further complications reported.
 
Additional Manufacturer Narrative
H3: product analysis of part # 3990003 ; lot # nm13h015 analysis summary: visual and optical inspection confirmed a thread of the inner threads of the inserter has been chipped.This type of damage is consistent with the engagement process of the short threaded shaft when inserted into the short inserter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSTONE CONTROL¿ SPINAL SYSTEM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12588821
Report Number1030489-2021-01242
Device Sequence Number4079961
Product Code LXH
UDI-Device Identifier00643169091436
UDI-Public00643169091436
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120368
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3990003
Device Catalogue Number3990003
Device Lot NumberNM13H015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2021
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date10/07/2021
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/04/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age88 YR
Patient SexFemale
Patient Weight47 KG
-
-