MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK TIBIAL ONLAY INSERT-SZ 4-10MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK TIBIAL ONLAY INSERT-SZ 4-10MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Model Number 180704-3
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032)
Date of Event 06/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial report submitted under mfr# 0002249697-2021-01288.Upon additional information received, it was confirmed the device belongs to a different registration number.The following devices were also listed in this report: mck femoral-lm-rl-sz 2; cat # 180502; lot # 545795-m.Mck tibial baseplate-lm/rl-sz 4; cat # 180604; lot # ms00060319.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Reported event: an event regarding rom involving a mako insert was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: materials reviewed: 06/11/2020: stryker pi report.(b)(6) 2020: intraoperative record.Left tka.Pain cocktail and vanco powder.Cemented with palacos r.Restoris mck onlay uni 4 x10mm poly, #2 lm/rl femoral, #4 tibial baseplate.(b)(6) 2020: operative note.Left uni medial knee.Mako.(b)(6) yo woman.Recut due to balance issues.Too tight, downsized femur to a #2.No complications.(b)(6) 2020: h and p.Ongoing pain in knee.Stable medical history.Stable ligaments, mild decrease in strength.Motion 0-1000.X-rays bone on bone in varus.Failed conservative rx.Kss 50 pain and function.Confirmation: the event cannot be confirmed as there is no documentation of the manipulation event.Root cause: a root cause cannot be confirmed as there is no documentation of the reported event.Product history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Information received from study site indicates "knee manipulation under anesthesia.".
 
Additional Manufacturer Narrative
Reported event an event regarding rom involving a mako insert was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: materials reviewed: 06/11/2020: stryker pi report.06/11/2020: intraoperative record.Left tka.Pain cocktail and vanco powder.Cemented with palacos r.Restoris mck onlay uni 4 x10mm poly, #2 lm/rl femoral, #4 tibial baseplate.06/11/2020: operative note.Left uni medial knee.Mako.62 yo woman.Recut due to balance issues.Too tight, downsized femur to a #2.No complications.06/11/2020: h and p.Ongoing pain in knee.Stable medical history.Stable ligaments, mild decrease in strength.Motion 0-1000.X-rays bone on bone in varus.Failed conservative rx.Kss 50 pain and function.Confirmation: the event cannot be confirmed as there is no documentation of the manipulation event.Root cause: a root cause cannot be confirmed as there is no documentation of the reported event.-product history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Information received from study site indicates "knee manipulation under anesthesia.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK TIBIAL ONLAY INSERT-SZ 4-10MM
Common Device Name
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key12590477
Report Number3005985723-2021-00167
Device Sequence Number3370136
Product Code NPJ
UDI-Device Identifier00848486000905
UDI-Public00848486000905
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K090763
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/23/2021
Device Model Number180704-3
Device Catalogue Number180704-3
Device Lot Number12180516-1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/15/2021
Supplement Date Received by Manufacturer09/15/2021
Initial Report FDA Received Date10/07/2021
Supplement Report FDA Received Date10/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age62 YR
-
-