Model Number 8888145017
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, pre-operative flushing of the catheter, some obstruction was observed in the arterial lumen.Both lumens were flushed, the venous lumen did not have any issues and flushing was not successfully done in arterial lumen.The catheter was discarded and a new catheter kit was opened.After the successful flushing of the new catheter, it was successfully inserted to the patient and subsequent dialysis was completed.The catheter was not repaired, there was no leak, betadine and chlorhexidine was used as the cleaning agent on the device.Tego was not utilized, there was no luer adapter issue and there was no patient symptoms or complications associated with this event.There was no reported patient injury.
 
Event or Problem Description
According to the reporter, pre-operative flushing of the catheter, some obstruction was observed in the arterial lumen.Both lumens were flushed, the venous lumen did not have any issues and flushing was not successfully done in arterial lumen.A new catheter kit was opened.After the successful flushing of the new catheter, it was successfully inserted to the patient and subsequent dialysis was completed.There was no blood loss and blood transfusion was not required.Nothing unusual was observed on the device prior to use, there was no other product utilized with the device, there were no other defects/damages found on the product, there was no damage to the device's external packaging and box.Betadine was used to treat the insertion site prior to product placement and there was no medical intervention done to the patient.The catheter was not repaired, there was no leak, betadine and chlorhexidine was used as the cleaning agent on the device.Tego was not utilized, there was no luer adapter issue and there were no patient symptoms or complications associated with this event.There was no reported patient injury.
 
Additional Manufacturer Narrative
Additional information: b5, d9, g1 (mfr contact first name, last name, email, phone number), g3, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device was occluded.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS   20101
Manufacturer Contact
clay chandler
5920 longbow drive
boulder, CO 80301
3035306409
MDR Report Key12590790
Report Number3009211636-2021-00283
Device Sequence Number3431480
Product Code MSD
UDI-Device Identifier10884521013193
UDI-Public10884521013193
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K111372
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2025
Device Model Number8888145017
Device Catalogue Number8888145017
Device Lot Number2008400130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2021
Supplement Date Received by Manufacturer01/31/2022
02/23/2022
Initial Report FDA Received Date10/07/2021
Supplement Report FDA Received Date02/17/2022
02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-