• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL LINEAR STAPLER; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL LINEAR STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Model Number TX30G
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # 225a11.Device analysis: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tx30g device arrived with no apparent damaged and with a reload loaded on the device.The reload was received fully loaded with staples.The device was tested for functionality with the returned reload and it fired and formed all the staples as intended.The device fired without any difficulties, the staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: "squeeze the closing trigger until a click is heard.The instrument is in an intermediate position, the pin is fully seated in the anvil capturing the tissue, and the jaws are partially open.Reposition tissue within the instrument if desired.Squeeze the closing trigger and the handle fully together until a second click is heard.The closing trigger is now latched to the handle and the jaws are clamped onto the tissue, which is ready to be stapled.The firing trigger will simultaneously move down to the ready-to-fire position.Before firing, check to ensure the retaining pin is seated in the anvil.If the pin is not properly positioned, staples may not form properly, which may result in leakage or disruption of the staple line.Fire the instrument by pulling the firing trigger back completely against the closing trigger until a click is heard, signifying the staples are fully formed.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
 
Event or Problem Description
It was reported that during an anterior resection, when using the tx30g to staple across the bowel it misfired.The stapler was closed.They waited 15 seconds and pulled the firing handle, and it did no fire staples.Opened another tx30g and fired with no issues.Patient was not harmed.There were no patient consequences.The patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
MDR Report Key12590791
Report Number3005075853-2021-06032
Device Sequence Number3292554
Product Code GDW
UDI-Device Identifier10705036003328
UDI-Public10705036003328
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2026
Device Model NumberTX30G
Device Catalogue NumberTX30G
Device Lot Number242A55
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/16/2021
Initial Report FDA Received Date10/07/2021
Patient Sequence Number1
-
-