MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLSTENT ENDOPROSTHESIS; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLSTENT ENDOPROSTHESIS; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 26280
Medical Device Problem Codes Positioning Failure (1158); Difficult to Remove (1528); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter facility name: (b)(6).
 
Event or Problem Description
It was reported that removal difficulties and stent damage occurred.The target lesion was located in the iliac vein.A 16x90/9fr wallstent endoprosthesis was advanced for treatment along with the guidewire.However, the stent was stuck during deployment and was unable to be deployed nor withdrawn.The physician pulled out the delivery shaft, it was noted that the stent was elongated and deformed inside the patient.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
E1: initial reporter facility name: (b)(6) hospital.Device evaluated by mfr: a 16x90/9fr uni plus 75cm was received for analysis.A visual examination found that the stent of the device had been deployed from the delivery system.The deployed stent was not returned for analysis.A visual inspection of the stent impression and stent cups identified no issues.The imprinted stent impression was clear on the stent holder.A visual and tactile examination identified no issues with the tip of the device.A visual and tactile examination identified no kinks or damage along the length of the device.
 
Event or Problem Description
It was reported that removal difficulties and stent damage occurred.The target lesion was located in the iliac vein.A 16x90/9fr wallstent endoprosthesis was advanced for treatment along with the guidewire.However, the stent was stuck during deployment and was unable to be deployed nor withdrawn.The physician pulled out the delivery shaft, it was noted that the stent was elongated and deformed inside the patient.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENDOPROSTHESIS
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12591357
Report Number2134265-2021-12552
Device Sequence Number11857205
Product Code MAF
UDI-Device Identifier08714729204053
UDI-Public08714729204053
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152842
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/18/2023
Device Model Number26280
Device Catalogue Number26280
Device Lot Number0026651531
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/27/2021
Supplement Date Received by Manufacturer12/02/2021
Initial Report FDA Received Date10/07/2021
Supplement Report FDA Received Date12/16/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age35 YR
Patient SexMale
Patient Weight56 KG
-
-