| Model Number |
7580 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a shaft detachment occurred.This 2.50mm x 12mm maverick balloon catheter was selected for a percutaneous coronary intervention.During removal from the lesion difficulties were encountered and the distal end of the balloon became stuck inside the guide catheter and broke.If was noted that the balloon was deflated prior to removal.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft detachment occurred.This 2.50mm x 12mm maverick balloon catheter was selected for a percutaneous coronary intervention.During removal from the lesion difficulties were encountered and the distal end of the balloon became stuck inside the guide catheter and broke.If was noted that the balloon was deflated prior to removal.No patient complications were reported.It was further reported that the lesion was 90-99% stenosed, without a significant bend and located in the first marginal.The lesion was treated via shockwave treatment but the convey guide catheter was not stable so they selected this maverick balloon catheter which was put over another manufacturers guidewire for back up support.During withdrawal through the forearm significant resistance was met and the maverick became stuck inside the middle of the convery guide catheter, they attempted to remove the device by using another manufactures balloon catheter, without success.They removed all the devices from the patient, rinsed the guide catheter and the last piece of the maverick device was removed from the patient.The device was completely removed from the patient and the patient status is ok.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the returned product consisted of a maverick 2 mr balloon catheter.The device was visually and microscopically examined.There was contrast and blood in the inflation lumen and balloon.There was blood in the guidewire lumen, and the shaft had numerous kinks.27.5cm from the tip of the device, the shaft was separated.The ends of the separation were jagged indicating the shaft was torn.The inflation lumen on the proximal end of the separation was stretched down for a total length of 32.7cm.The shaft of the distal portion of the separation was stretched including both the inflation lumen and guidewire.The balloon was loosely folded.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft detachment occurred.This 2.50mm x 12mm maverick balloon catheter was selected for a percutaneous coronary intervention.During removal from the lesion difficulties were encountered and the distal end of the balloon became stuck inside the guide catheter and broke.If was noted that the balloon was deflated prior to removal.No patient complications were reported.
|
| |
|