Medtronic received a report that three pipelines would not open distally.The patient was undergoing surgery for treatment of a fusiform, unruptured aneurysm in the right internal carotid superclinoid artery with a max diameter of 8mm and a 2mm neck diameter.The landing zone diameter was 3.74mm on the distal end and 5.1mm at the proximal end.It was noted the patient's vessel tortuosity was severe.Dapt (dual antiplatelet treatment) was administered and the pru level was administered.Angiographic results were the same as pre-procedure.Medical history included lung cancer ( 1 yr clear), and diverticulitis.It was reported that the pipeline would not open distally.The pipeline was not positioned in a bend.Less than 50% of the pipeline had been deployed when it failed to open.It was resheathed more than 2 times.Additional steps included multiple manipulations, recapture, wagging, and pulling the device back.The pipeline was reshaethed and removed with the microcatheter.The doctor is an experienced pipeline user.This was tortuous anatomy.This happened three times where the pipelines failed to open at all distally.The doctor aborted the procedure. the pipeline was not used for an indication that is not approved (off-label) and the pipeline and any accessory devices prepared as indicated in the ifu. ancillary devices include a phenom 27 microcatheter.
|
|
H3: the pipeline flex was returned within the phenom 27 catheter.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The distal and proximal ends of the pipeline flex braid were fully opened and frayed.No bend found on the pusher.No damages were found with the tip coil, distal marker, re-sheathing marker or with the proximal bumper.No other anomalies were observed.The pipeline flex was pushed out from the catheter lumen with no issue.Based on the returned device, the pipeline flex was not confirmed to have failure to open at the distal end as the distal and proximal ends of the pipeline flex braid were found fully opened and frayed.The damage to the braid on the ends of the pipeline flex is likely the results of the physician re-sheathing the device more than recommended two times.It is possible that the severe vessel tortuosity may have contributed to the failure to open issue.Per our instructions for use (ifu): ¿begin to deliver the device using a combination of unsheathing the pipeline flex embolization device and pushing delivery wire simultaneously.After the distal end of the pipeline flex embolization device has successfully expanded, deploy the remainder of pipeline flex embolization device by pushing the delivery wire and/or unsheathing the pipeline flex embolization device.Resheathing and/or manipulation of the micro catheter, by locking down the delivery wire and moving both as a system, may facilitate expansion of the pipeline flex embolization device.The system is designed to allow for 2 full cycles of re-sheathing of the pipeline flex embolization device.Re-sheathing the pipeline flex embolization device more than 2 full cycles may cause damage to the distal or proximal ends of the braid.Do not use in pati ents in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|