ARTHROCARE CORP. STARVAC 90 ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ASC4251-01
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the packing seal of the starvac 90 icw packaged was broken.No case reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
The reported device was received for evaluation.A visual inspection found the orange tamper seal is intact.The label that seals the opposite end of the package was torn.The sterile package was not damaged.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.A review of the packaging process found that tamper seal and label are applied to the packaging.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There was no way to determine if the device contributed to the reported event.The complaint was confirmed, but the root cause could not be determined.Factors that could have contributed to the reported event include damage caused by shipping or handling.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARVAC 90 ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12594325
Report Number3006524618-2021-00959
Device Sequence Number3292561
Product Code GEI
UDI-Device Identifier00817470001716
UDI-Public00817470001716
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberASC4251-01
Device Catalogue NumberASC4251-01
Device Lot Number2071159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2021
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date10/07/2021
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-