| Model Number |
ASC4251-01 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the packing seal of the starvac 90 icw packaged was broken.No case reported; therefore, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was received for evaluation.A visual inspection found the orange tamper seal is intact.The label that seals the opposite end of the package was torn.The sterile package was not damaged.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.A review of the packaging process found that tamper seal and label are applied to the packaging.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There was no way to determine if the device contributed to the reported event.The complaint was confirmed, but the root cause could not be determined.Factors that could have contributed to the reported event include damage caused by shipping or handling.No containment or corrective actions are recommended at this time.
|
| |
|