• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. T-CONN W/LL & 1 VLV PORT; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. T-CONN W/LL & 1 VLV PORT; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 20041E
Medical Device Problem Codes Leak/Splash (1354); Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported t-conn w/ll & 1 vlv port was damaged and leaked.The following information was provided by the initial reporter: "the hub connected with syringe broken.".
 
Event or Problem Description
It was reported t-conn w/ll & 1 vlv port was damaged and leaked.The following information was provided by the initial reporter: "the hub connected with syringe broken".
 
Additional Manufacturer Narrative
H6: investigation summary: a 20041e product was not available for investigation, however customer also provided a photograph of affected sample analysis of the photograph confirmed the customer's experience with the female luer observed to be cracked.Further information provided by the customer confirmed that the leakage occurred between female luer and syringe at the beginning of infusion.The details of this feedback were forwarded to the manufacturing site for investigation.A review of the production records for lot 21016355 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.Previous complaints of this nature have been caused or contributed by a combination of different factors, including over-torque of the component during connection with the male luer, use of a male luer that is not compliant to iso standards, or prolonged use of substances that are aggressive to plastics.In this instance the connecting product in use at the time of the customer's experience was not available for investigation and therefore it has not been possible to confirm if this may have contributed to the customer's experience.The quality team at the manufacturing site has been informed of this report in order to be aware of the reported feedback during future production of this product.A review of the customer feedback database indicates that this is an isolated occurrence with no other similar report against the 20041e set in the past 12 months.H3 other text : see h10.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 11/10/2021.H.6.Investigation: one 20041e sample was received without packaging, however the customer indicated that the complaint sample was from lot 21016355.A connecting luer slip syringe was also received and residual blood was present in the line.A visual inspection confirmed the customer's experience with the female luer observed to be cracked; leakage was observed from the crack during a flush through of the set.A review of the production records for lot 21016355 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.Previous complaints of this nature have been caused or contributed by a combination of different factors, including over-torque of the component during connection with the male luer, use of a male luer that is not compliant to iso standards, or prolonged use of substances that are aggressive to plastics.In this instance based on the information available it has not been possible to confirm which of these factors may have contributed to the customer's experience.H3 other text : see h.10.
 
Event or Problem Description
It was reported t-conn w/ll & 1 vlv port was damaged and leaked.The following information was provided by the initial reporter: "the hub connected with syringe broken".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-CONN W/LL & 1 VLV PORT
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12594991
Report Number9616066-2021-52185
Device Sequence Number14508347
Product Code FPA
UDI-Device Identifier27613203021665
UDI-Public27613203021665
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/28/2024
Device Model Number20041E
Device Catalogue Number20041E
Device Lot Number21016355
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2021
Supplement Date Received by Manufacturer10/14/2021
11/11/2021
Initial Report FDA Received Date10/07/2021
Supplement Report FDA Received Date11/08/2021
12/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-