| Model Number |
180308HC-S-V-A2 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The pusher and introducer were the only components returned for analysis.The investigation of the returned pusher found the implant to be separated from the pusher at the attachment bond section.The attachment monofilament profiles a tensile break at the attachment bond which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage or when this occurred, but the damage is consistent with the device experiencing forces over specification.
|
| |
|
Event or Problem Description
|
|
It was reported that an embolization coil implant was missing from the delivery pusher.There was no patient injury or intervention.The patient's condition is reported to be "good.".
|
| |
|
Search Alerts/Recalls
|