Unomedical reference number (b)(4).Event occurred in the united states.On (b)(6) 2021, it was reported that the patient's infusion set's tubing came apart at the site location.The site location was patient's right thigh, and the pump was at right side of patient's waist.Moreover, the infusion had been used for the first day.Reportedly, the infusions were not stored, or used, in a place where they might have been exposed to extreme temperatures and humidity.There was no stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.Currently, patient's blood glucose level was 286 mg/dl.No further information available.
|