• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORCYM S.R.L. CARBOMEDICS TOP HAT MECHANICAL HEART VALVE; MECHANICAL HEART VALVE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORCYM S.R.L. CARBOMEDICS TOP HAT MECHANICAL HEART VALVE; MECHANICAL HEART VALVE PROSTHESIS Back to Search Results
Model Number CPHV
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/08/2021
Type of Reportable Event Death
Additional Manufacturer Narrative
A complete manufacturing and material records review for the device has been performed.The results confirmed that the device satisfied all material, visual and performance standards required at the time of manufacture and release.Since the device was not returned to the manufacturer (not explanted), no further investigation is possible at this time.Based on the available information, it is not possible to establish a definitive root cause for the reported event.However, from the document review performed, no manufacturing deficiencies were identified.It is reported that the device model s5 was checked during the emergency ultrasound that showed an inadequate opening motion.Since however the patient had a sudden blood loss, it is not possible to judge whether the inadequate opening was a cause or a result of the reported event.At the end of the surgical procedure, a good valve functionality was detected.Given that no further investigation on the device was possible, and since no medical judgment was received on the event, the root cause remains unknown at this time.Should the manufacturer receive additional information in the future, a follow up report will be provided.
 
Event or Problem Description
On (b)(6) 2021 the patient underwent aortic and mitral valve replacement with corcym devices due to lesions on both valves.The patient received an optiform size 27 in mitral position (ref (b)(4)) and a top hat size 21 in aortic position.The devices were safely implanted during the surgery and patient did well during surgery.There was no abnormality or malfunction noted on the valve during surgery.The valves were tested after implantation and confirmed the position and opening and closing function was good.After the operation, the heart rate and blood pressure were stable, so the patient was then transferred to icu for observation.Two hours post-operation, the patient had a sudden loss of blood pressure (blood pressure 0), but the heartbeat remained.Then, the patient died.The emergency ultrasound showed inadequate opening and closing motion for s5 valve.Since no further investigation on the patient was possible, the site could not assess the relationship between the devices and the event, nor if the devices malfunctioned or not.No further information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBOMEDICS TOP HAT MECHANICAL HEART VALVE
Common Device Name
MECHANICAL HEART VALVE PROSTHESIS
Manufacturer (Section D)
CORCYM S.R.L.
strada crescentino
saluggia, vercelli 13040
IT  13040
MDR Report Key12598308
Report Number3005687633-2021-00161
Device Sequence Number19729387
Product Code LWQ
UDI-Device Identifier08022057012906
UDI-Public(01)08022057012906(240)S5-021(17)260321
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P900060
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/21/2026
Device Model NumberCPHV
Device Catalogue NumberS5-021
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2021
Initial Report FDA Received Date10/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age67 YR
-
-