| Model Number |
70914-70 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
Loss of consciousness (2418)
|
| Date of Event |
08/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint, and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the freestyle freedom lite meter and freestyle strips were reviewed, and the dhrs showed the freestyle lite meter and freestyle strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed within specification.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
The wife of customer reported a high reading issue with the adc blood glucose meter, and reported that the customer "collapsed".However, no treatment was required.The caller stated she was unsure if this event was due to reported issue or because customer just had pancreatic cancer surgery.There was no report of death or permanent injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported meter and test strips have been returned and investigated.Visual inspection has been performed on the returned meter and no issues were observed.Meter powered on with button press and strip insertion.Meter did not display an error message.Meter advanced to the apply sample screen.Control solution testing has been performed with returned meter and returned strips and all results were within range specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
The wife of customer reported a high reading issue with the adc blood glucose meter, and reported that the customer "collapsed".However, no treatment was required.The caller stated she was unsure if this event was due to reported issue or because customer just had pancreatic cancer surgery.There was no report of death or permanent injury associated with this event.
|
| |
|
Search Alerts/Recalls
|