• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-15-T; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-15-T; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 133S-MX-15-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.Device evaluation: yellow particles device inner surface and one curved opening.A microscopic analysis and leak test were performed which identified three openings striated.Based on the device analysis the final assessment is: three openings striated in the side suture tap/radius assessed as surgical damage.
 
Event or Problem Description
Healthcare professional reported right side deflation.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATRELLE TE SMOOTH 133S-MX-15-T
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
MDR Report Key12599143
Report Number9617229-2021-54129
Device Sequence Number14251438
Product Code LCJ
UDI-Device Identifier10888628044210
UDI-Public10888628044210
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K182054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/16/2024
Device Model Number133S-MX-15-T
Device Catalogue Number133S-MX-15-T
Device Lot Number3381270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/09/2021
Initial Report FDA Received Date10/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-