DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/16/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had a gas leak.Therapy was delayed until another unit was procured.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.Fse could not duplicate the problem.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted when our investigation is completed.(b)(6).
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, g7, h2, h4, h6, h10.
|
|