DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/16/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during a scheduled preventative maintenance (pm) performed by a getinge field service engineer (fse) that the cs300 intra-aortic balloon pump (iabp) had a premature discharge of the battery.There was no patient involvement thus no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device(10) fse replaced batteries.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.A supplemental report will be submitted when our investigation is completed.The initial reporter named is a getinge employee whose contact details differ from that of the event site.A contact name at the event site is: (b)(64).The full name of the event site was shortened due to field character limit; the full name is (b)(6).
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, g7, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person).
|
|
Event Description
|
N/a.
|
|