• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC BOLT FOR FEMORAL NECK SYSTEM 80MM LENGTH-STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC BOLT FOR FEMORAL NECK SYSTEM 80MM LENGTH-STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 04.168.280S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported on (b)(6) 2021, the patient underwent removal of femoral neck system (fns) due to nonunion and patient sustained a fall.There was a implant breakage.It took approximately (40) minutes to get the hardware out.Eventually everything was removed successfully.The patient is doing well at this point.This report is for (1) bolt for femoral neck system 80mm length-sterile.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable lot # provided is not a valid number, therefore, the dhr could not be completed.If device is returned or lot number can be confirmed, the dhr will be revisited.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOLT FOR FEMORAL NECK SYSTEM 80MM LENGTH-STERILE
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen 2540
SZ   2540
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key12601403
Report Number2939274-2021-05986
Device Sequence Number4112506
Product Code KTT
UDI-Device Identifier10886982274724
UDI-Public(01)10886982274724
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K172872
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.168.280S
Device Catalogue Number04.168.280S
Device Lot Number50P9363
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2021
Supplement Date Received by Manufacturer11/03/2021
Initial Report FDA Received Date10/08/2021
Supplement Report FDA Received Date11/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - NAIL HEAD ELEMENTS: FNS ANTIROTATION; UNK - PLATES: FNS; UNK - SCREWS: LOCKING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 YR
Patient SexUnknown
Patient Weight149 KG
-
-