| Model Number |
18-3356 |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
09/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No device deficiency was reported.Cardiac tamponade is a known potential adverse event of catheter ablation procedures disclosed in product labeling.The blood pressure drop was noted coincident with the positioning of another manufacturer's mapping catheter and sheath in the right superior pulmonary vein.The deflectable sheath was then moved to the left inferior pulmonary vein and further blood pressure decrease was noted.The physician did not conclusively exclude the deflectable sheath as a possible contributing cause of the cardiac tamponade.It was also reported the ablation catheter tip was completely withdrawn into the deflectable sheath.This is contrary to product labeling in which the soft distal tip of the catheter is recommended to be extended from the sheath during sheath movemement.It cannot be determined if this deviation from the instructions from use contributed to this adverse event.
|
| |
|
Event or Problem Description
|
|
During a pulmonary vein isolation procedure to treat atrial fibrillation, a drop in blood pressure occurred and cardiac tamponade was diagnosed with echocardiography.The cardiac tamponade was successfully treated by pericardiocentesis.The physician believes the cause of the cardiac tamponade was an electrophysiology catheter and sheath, made by a different manufacturer, which were used to check for potentials in the right superior pulmonary vein.However, the physician could not exclude the deflectable sheath as a possible cause of the cardiac tamponade.
|
| |
|
Search Alerts/Recalls
|