• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TRUEPASS NEEDLE SINGLE PACK STERILE BX 5; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TRUEPASS NEEDLE SINGLE PACK STERILE BX 5; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72203793
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Pain (1994); Foreign Body In Patient (2687)
Date of Event 09/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that during shoulder surgery, internal to patient, the truepass needle single pack broke, the fragments that broke off are lost in the patient's shoulder but visible on x-rays.The procedure was successfully completed with a significant delay using a backup device.The patient is reported to have residual pain.
 
Additional Manufacturer Narrative
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Insufficient product identification information was provided, and thus, a manufacturing record review could not be conducted.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately and there were no indications to suggest the anticipated risk is not adequate.This case reports the breakage of the truepass needle, which per case details, was retained with in the patient.No clinically relevant supporting documentation was provided for inclusion in this medical investigation.Based on the limited information provided the root cause for the reported breakage could not be determined.The truepass needle tip and shaft is made of nitinol which is a biocompatible material however, since the needle is part of a surgical device it is not approved for implantation.The patient impact beyond possible corrosion, local irritation/discomfort, and/or migration of possibly retained non-implantable foreign body fragments cannot be determined.No further medical assessment can be rendered at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Event or Problem Description
It was reported that during shoulder surgery, the truepass needle broke inside of the patient.The fragments that broke off are lost in the patient's shoulder but visible on x-rays.The procedure was successfully completed with a surgical delay greater than 30 minutes using a backup device.The patient is reported to have residual pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEPASS NEEDLE SINGLE PACK STERILE BX 5
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12603028
Report Number1219602-2021-02246
Device Sequence Number11097030
Product Code NBH
UDI-Device Identifier00885554030433
UDI-Public00885554030433
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203793
Device Catalogue Number72203793
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2021
Supplement Date Received by Manufacturer12/21/2021
Initial Report FDA Received Date10/10/2021
Supplement Report FDA Received Date12/24/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-