| Model Number |
072217 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a hair found mixed in the package of the bard channel drain.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The device has not met specifications.The product was not used for patient treatment.The product had caused the reported failure.Visual evaluation of the returned sample noted one unopened (with original packaging), channel drain.Visual inspection of the sample noted a piece of hair found in the sealed packaging.A potential root cause for this failure could be defective / contaminated components from supplier.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that there was hair found mixed in the package of bard channel drain.
|
| |
|
Search Alerts/Recalls
|