| Model Number |
86623 |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
A ls pro recording system was selected for use.During preparation for an ablation procedure, patient id was entered into the system and the patient window was closed.An error message 'you will not be able to use any features that require the amplifier.Contact boston scientific technical support if you need assistance' was observed."cancel" or "ok" was pressed the ls pro application was closed.The workstation was rebooted and the application was relaunched, however, error was not resolved.The amplifier was powered on properly and no errors were observed during the self-test and green flashing "system ok" led was seen.Ethernet cable connection was checked with no issues.The procedure was cancelled at this point.
|
| |
|