• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LABSYSTEM PRO; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LABSYSTEM PRO; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 86623
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event or Problem Description
A ls pro recording system was selected for use.During preparation for an ablation procedure, patient id was entered into the system and the patient window was closed.An error message 'you will not be able to use any features that require the amplifier.Contact boston scientific technical support if you need assistance' was observed."cancel" or "ok" was pressed the ls pro application was closed.The workstation was rebooted and the application was relaunched, however, error was not resolved.The amplifier was powered on properly and no errors were observed during the self-test and green flashing "system ok" led was seen.Ethernet cable connection was checked with no issues.The procedure was cancelled at this point.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LABSYSTEM PRO
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key12605560
Report Number2134265-2021-12694
Device Sequence Number1415268
Product Code DQK
UDI-Device Identifier08714729935872
UDI-Public08714729935872
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K141185
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86623
Device Catalogue Number86623
Device Lot Number0001972405
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2021
Initial Report FDA Received Date10/11/2021
Patient Sequence Number1
-
-