• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BL, Ø 3.3MM NC, SLACT. 12MM, RXD, GUIDED; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BL, Ø 3.3MM NC, SLACT. 12MM, RXD, GUIDED; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number SLActive Roxolid BL
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Fibrosis (3167)
Date of Event 06/03/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2021 in ada 4.Details of surgery: implant surface not completely covered with bone.On (b)(6) 2021, non-osseointegration was verified.The device was forwarded to the manufacturer.At the event the patient experienced: infection, pain and mobility.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BL, Ø 3.3MM NC, SLACT. 12MM, RXD, GUIDED
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
MDR Report Key12607132
Report Number0001222315-2021-17139
Device Sequence Number19162666
Product Code DZE
UDI-Device Identifier07630031706430
UDI-Public07630031706430
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 10/11/2021
Report Date (Section F) 09/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/03/2024
Device Model NumberSLActive Roxolid BL
Device Catalogue Number021.2212G
Device Lot NumberWY396
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2021
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/11/2021
Device Age23 MO
Event Location Other
Date Report to Manufacturer09/16/2021
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-