COOK ENDOSCOPY FUSION PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
| Model Number |
G34693 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook fusion® pre-loaded with acrobat wire guide.It was initially reported that the physician cannulated initially and performed a sphincterotomy.Then, the sphincterotome was removed.They had difficulty withdrawing it from the wire lock.They were able to get it out.At that point, they passed an extraction balloon though the scope and performed balloon sweeps.After that point, the physician then passed the sphincterotome through the scope and had trouble again passing it through the wire lock.Once the tome was passed through and visible in the duodenum, the cutting wire was broken.They removed it and successfully completed the procedure with another device of the same type.There was no reportable information at this time.Our evaluation of the returned device determined that the cutting wire securing component (anchor) has separated from the catheter (w/ detachment) resulting in tip will not bow.[subject of report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # k172288.Investigation evaluation: our laboratory evaluation of the product said to be involved determined the cutting wire securing component located near the distal end of the sphincterotome has disconnected from the catheter.However, the cutting wire is intact and remains securely attached to the sphincterotome at the proximal end.A section of the cutting wire securing component has broken and detached from the device.The broken section is estimated to be 3.0mm in length and 0.5mm in diameter.The broken section was not included in the return.There was liquid observed in the catheter and the wire guide lumen was partially utilized.The associated wire guide also returned and was noted to be slightly bend at the distal end after use.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user: "do not over flex or bow tip beyond 90 degrees, as this may damage or cause cutting wire to break." other factors that can contribute to separation of the cutting wire securing component and the catheter include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all - fusion® pre-loaded with acrobat wire guide are subjected to a visual inspection to ensure device integrity.The visual inspection ensures the product is free of kinks.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Search Alerts/Recalls
|
|
|