| Model Number |
TN,DXH 520 CP HEMATOLOGY INSTRUMENT |
| Medical Device Problem Code |
Patient Data Problem (3197)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer technical support (cts) specialist discussed the maintenance of the instrument and the customer said that the instrument was due for monthly bleaching.The customer was advised to bleach and clean the wbc filter and to call back if the issue was not resolved.The customer agreed to do so and has not called back.Bec internal identifier case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that their dxh520 cp hematology instrument had generated erroneous high platelet (plt) results on patient samples.Erroneous results were not reported outside of the laboratory.There was no report of serious injury or impact to patient treatment related to this event.Printouts were requested multiple times but not provided by the customer.
|
| |
|