| Model Number |
TN,DXH 520 CP HEMATOLOGY INSTRUMENT |
| Medical Device Problem Code |
Patient Data Problem (3197)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) observed that the customer's two dxh520 (system id (b)(4) )have same device id (20) in the lis setup screen.The lis uses the instrument sequence number as the primary identifier to chart patient results into electronic record.There was one instance were both instruments had the same sequence number (same patient id) that resulted in a charting error.The customer was able to detect error and correct charting issue.The fse changed the device id on system id (b)(4) from id 20 to id 21 to correct duplicate device id issue.The fse also recommended to the lis vendor to change primary identifier from the sequence number to the patient id as best practice.Bec internal identifier (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported there was one instance were both their dxh520 cp hematology instruments had the same sequence number (duplicate patient id) that resulted in a charting error.Erroneous results were reported outside the laboratory, however, there was no reported change or effect on patient treatment.There was neither death or serious injury associated with the event.The customer detected and corrected the results of the charting issue.
|
| |
|