SMITH & NEPHEW, INC. FLEXIBLE SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
71362290 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a reflection flexible screwdriver is broken.As this was noticed in a non-surgical environment, there was not any patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned flexible screwdriver confirms the device is chipped on the edges.This device also has extensive wear usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).Further review by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The event states that the flexible screwdriver was found damaged during inspection without patient involvement.Users are instructed to inspect all instruments before use and not to use the device if any damage is found.Device damage that is identified during the inspection process will not cause or contribute to a serious injury or death, nor would it be likely to do so in the event of recurrence as the inspection process itself mitigates the risk of any potential harm occurring.Therefore, it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If additional details confirm the occurrence of a reportable event, we will update our records accordingly and submit a subsequent report detailing both the event and our complete investigations.
|
| |
|