MAUDE Adverse Event Report: SYNTHES GMBH VERTECEM V+ SYRINGE KIT; POLYMETHYLMETHACRYLATE BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH VERTECEM V+ SYRINGE KIT; POLYMETHYLMETHACRYLATE BONE CEMENT Back to Search Results
Catalog Number 03.702.215S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that this was a percutaneous vertebroplasty (l1) for treating compression fracture due to osteoporosis on (b)(6), 2021.After the cement was mixed up, the staff checked the mixed cement to confirm it was ready for use.The air in the cement mixing device was squeezed out.Next, the 2-way connector was mounted however, the cement did not come out.It was confirmed that the cement on the mixer side flowed but nothing changed on the syringe side of the connector.The black rubber ring located inside the cement mixers cap came off during troubleshooting.The procedure was completed with replacements and required less than 30-minute surgical delay.No further information is available.This report is for one (1) vertecem v+ syringe kit.This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6- visual analysis of the photo revealed that vertecem v+ syringe kit cement has been solidified in the stopcock and found that the ring-shaped rubber packing of the lid has fallen off.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for vertecem v+ syringe kit.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> part: 03.702.215s.Lot: 1012516.Manufacturing site: selzach.Supplier: symatese device.Release to warehouse date: 03 february 2021.Expiration date: 01 february 2026.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: investigation summary: the product was returned to us cq for evaluation.The one-way stop cock was only returned from vertecem v+ syringe kit.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that cement was hardened inside the stock cock.The ring-shaped rubber packing of the lid was received separately from the stock cock.However, it was snapped into the stock cock and fits perfectly well.No other issues were identified with the device.The dimensional inspection was not performed for the vertecem v+ syringe kit as it's not pertaining to the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for the vertecem v+ syringe kit.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Corrected data: d9, h3, h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERTECEM V+ SYRINGE KIT
Common Device Name
POLYMETHYLMETHACRYLATE BONE CEMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12616042
Report Number8030965-2021-08610
Device Sequence Number4201280
Product Code NDN
UDI-Device Identifier07611819396074
UDI-Public(01)07611819396074
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2026
Device Catalogue Number03.702.215S
Device Lot Number1012516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/14/2021
Supplement Date Received by Manufacturer10/19/2021
10/19/2021
12/14/2021
Initial Report FDA Received Date10/12/2021
Supplement Report FDA Received Date10/19/2021
12/10/2021
12/14/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SISTEMA DE CEMENTO VERTECEM V+ . ESTERIL.
Patient SexUnknown
-
-