Catalog Number UNKNOWN
Device Problems Structural Problem (2506); Detachment of Device or Device Component (2907)
Patient Problems Perforation of Vessels (2135); Anxiety (2328); Unspecified Mental, Emotional or Behavioural Problem (4430)
Event Type  Injury  
Manufacturer Narrative
Occupation: non-healthcare professional.Investigation: the reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: fracture, vena cava (vc) perforation.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Catalog and lot numbers are unknown, however, the alleged celect is manufactured and inspected according to controls no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
The following information is alleged: the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2013 and struts of the filter have perforated the ivc.One strut abuts the l3 vertebral body; one strut has fractured proximally, protruding outside of the ivc lumen and another strut is missing.Hospital and medical records have been requested, but not yet provided.
 
Event Description
Patient allegedly received a filter via the right femoral vein due to deep vein thrombosis (dvt) of left lower extremity with pulmonary embolism (pe), despite anticoagulants.Patient is alleging vena cava perforation which is abutting an organ and device fracture.Patient notes and further alleges experiencing one filter strut "is abutting my l3 vertebral body.Further, one strut has fractured proximally, protruding outside of the ivc lumen.I live with the anxiety of having a filter that could fail further at any time, but that cannot be retrieved without a serious surgery.I am worried because my filter has fractured and perforated outside my vena cava abutting another organ", physical limitations, bipolar disorder, agoraphobia, and paranoia.
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: anxiety, worry, physical limitations, bipolar disorder, agoraphobia, paranoia.Unknown if the reported anxiety, worry, physical limitations, bipolar disorder, agoraphobia, and paranoia are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot numbers are unknown, however, the alleged celect is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
56868686
MDR Report Key12616626
MDR Text Key276423263
Report Number3002808486-2021-01865
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/15/2021
Initial Date FDA Received10/12/2021
Supplement Dates Manufacturer Received01/24/2022
Supplement Dates FDA Received02/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexMale
Patient Weight111 KG
-
-